The Universal Meter Reads ALL Heads: Power: Thermopiles, Photo Detectors and Pyroelectrics Energy: Thermopiles (in single shot mode), Photo Detectors and Pyroelectrics
Measure fJ Energy Levels. Thanks to a unique digital method for suppressing the noise on the lower ranges.
External Trigger. Synchronize you M-LINK to your pulsed laser or digital chopper.
Digital (USB) Output. Connect the M-LINK module directly to your PC.
Powerful LabView Software. Features include: Complete instrument controls: Range, Trigger, Wavelength, etc. Live display in J and J/cm2 or W and W/cm2 Full Statistics: Min, Max, Mean, Standard Deviation, RMS Stability, Repetition Rate, etc. Graphic Displays: Strip Chart, Histogram, Tuning Needle and more Data File Collection and Analysis
Universal Laser Power & Energy Meter 11M-LINK. This new PC-Based monitor is compatible with ALL types of detectors - including thermopiles, pyroelectrics and photo detectors - for both power and energy measurements. The device is available as a single channel unit that directly interfaces with a computer using a USB2.0 connection. The LabView software is included and comes with all the necessary features. The 11M-LINK also presents a unique digital technique of suppressing the noise, thereby extending the measurement range all the way down to the fJ level.
VERSATILE SOFTWARE FOR THE UNIVERSAL 11M-LINK What makes the 11M-LINK so universal is its compatibility with every detector type and model we make, and our smart software that recognizes the type of detector attached, and configures itself accordingly. Some of the basic software features include: Live Digital Reading, Full Statistics, Strip Chart, Histogram, Analog Tuning, Data Logging.
MEASURE POWER WITH A PHOTO DETECTOR If you need to measure low power levels, from pW to mW, then we recommend our 11PD100-Si detectors. In the software screen shown on the left, we have taken a data set working in the "STATS" display mode. We have set the batch size to 100 data points in the manual reset mode. You can see the live power (138 nW) and full complement of statistics: mean, max, min, RMS and PTP stability. In the bottom left hand corner you will note that a wavelength of 300 nm is displayed. This is where you will enter the wavelength of your laser and engage the wavelength correction factor.
MEASURE POWER WITH A THERMOPILE DETECTOR You can select any of our Thermal Detectors to measure your laser power from a few μW up to 30 kW. We used one of our most sensitive thermopile detectors, model 11XLP12-3S-H2, to generate the software screen shown on the left. We have selected the "SCOPE" mode, where you can view the live power reading (0.404 W), a bar graph and a strip chart while monitoring the power. This high level screen also provides access to range, trigger, auto scale, and many other monitor functions.
MEASURE ENERGY WITH A PYROELECTRIC DETECTOR You can select one of our Pyroelectric Detectors of the 11QE Series for energy measurements. We are looking at the "HISTOGRAM" screen, where you can continue to view the live measurement and a histogram that shows the energy distribution of your data set, along with a best-fit Gaussian curve. Note that you still have access to the instrument controls, like range, trigger, wavelength, etc.
MEASURE AT THE fJ LEVEL Take advantage of our proprietary pulse averaging, noise reduction techniques available with 11M-LINK. In the example shown at the right, we have captured a data set while running in the "AVERAGING" mode. The bars represent minimum (white) and maximum (blue) energy values. The strip chart is based on the average energy value. You get to select the number of "BINS" represented here. "Pulse Averaging" is available in the Statistics screen.
Compatible with all 11UP, 11XLP12 laser power detectors, 11PH photo detectors and 11QE laser energy detectors.
Specifications
Monitor
11M-LINK
Detector Types
ALL MODELS: Thermopiles, Pyroelectrics, Photo Detectors
Display
PC-Based
POWER METER SPECIFICATIONS
Power Range
4 pW to 30 kW
Resolution (Digital)
Current Scale/3000
Monitor Accuracy
±0.5 % ±2 digits
Statistics
Current Value, Max, Min, Average, Std Dev., RMS & PTP Stability, Time
ENERGY METER SPECIFICATIONS
Energy Range
30 fJ to 30 kJ
Resolution (Digital)
Current Scale/3000
Monitor Accuracy
1 % ± 2 digits (<1 kHz)
Software Trigger Level
0.1 to 99.9 %, 0.1 % resolution, default 2 %
Repetition Rate a
1 000 Hz
Real Time Data Transfer
1 000 Hz with time stamp, no missing point
Statistics
Current Value, Max, Min, Average, Std Dev., RMS & PTP Stability, Pulse #, Repetition Rate, Average Power
DETECTOR COMPATIBILITY
Thermopile
Average Power & Single Shot Energy
Pyroelectric
Pulse Energy & Average Power
Photo Detectors
Average Power & Pulse Energy
GENERAL SPECIFICATIONS
Digital Display
Computer Screen
Data Display
Real Time, Scope, Averaging, Statistics and Digital Tuning Needle
Serial Commands and Data Transfer
Via USB
Real Time Data Transfer Rate
1 000 Hz with time stamp, no missing point (for pyroelectrics only)
Analog Output
0-2 Volts, Full Scale, ± 2% (joulemeters) ± 4% (wattmeters)
Rising or Falling Edge External Trigger
4.5 to 10 V @ 20 mA, optically isolated
Dimensions
106W x 34H x 147D mm
Weight
0.424 kg
Specifications are subject to change without notice