Optical schematic |
Modified Czemy-Turner double monochromator supplemented with an optical matching unit providing either additive or subtractive dispersion |
Focal Length, mm -1st monochromator, collimating mirror -1st monochromator, focusing mirror -2nd monochromator, collimating mirror -2nd monochromator, focusing mirror |
130 142 142 130 |
Ports |
1 input, 1 output |
F/Number |
1 : 4.5 |
Entrance & Exit slits Slit width Slit height Parallelism Micrometer reading accuracy |
Micrometric adjustable from 0 to 2.0 mm 12 mm +/- 1 mkm +/- 1 mkm |
Intermediate slit Slit width Slit height Parallelism Micrometer reading accuracy |
Micrometric adjustable from 0 to 5.0 mm 12 mm +/- 1 mkm +/- 1 mkm |
GRATINGS * |
Interchangeable, 2 pairs supplied with the instrument |
Number of grooves per mm Grating size Blazing wavelength Usable wavelength range** Scanning wavelength range |
1200 25 x 25 x 8 mm 280 nm 190-650 nm 0 - 760 nm |
600 25 x 25 x 8 mm 600 nm 385-1300 nm 0 - 1300 nm |
WAVELENGTH Reciprocal Dispersion (average) of the first monochromator Wavelength accuracy Wavelength repeatability Wavelength step size Wavelength scan speed Wavelength resolution |
5.8 nm/mm +/- 0.15 nm +/- 0.05 nm 0.01 nm 19 nm/s 0.07 nm |
11.6 nm/mm +/- 0.25 nm +/- 0.1 nm 0.02 nm 38 nm/s 0.14 nm |
Dispersion Addition Mode Reciprocal dispersion (average) Width of the selected spectral range |
3.0 nm/mm 0.2 - 6 nm |
6.0 nm/mm 0.4- 12 nm |
Dispersion Subtraction Mode Reciprocal dispersion Width of the selected spectral range Width of 25 mkm-Entrance slit image -with 2.5 mm intermediate slit -with 5.0 mm intermediate slit |
0 0.2 - 40 nm
32 mkm 43 mkm |
0 0.4 - 80 nm
32 mkm 43 mkm |
Stray Light |
10-9 |
Computer Interface |
High Speed USB |
Electrical Service Requirement |
Single phase main 85 - 264 VAC; 47-63 Hz |
Dimensions and weight |
265 x 195 x 210 mm, 8 kg |